Rapidox 4100 Technical Characteristics

Analyzer dimension

250mm X 263mm X 150mm

Weight

3 kg

Power supply

110/120V or 220/240V, 50/60Hz; Max power consumption 4VA

Display

Multi-digit LCD - character height 12.7mm

Display ranges

E version - display range 0.01% to 100.0%. Suitable for samples containing mildly acidic gases; eg, carbon dioxide, hydrogen sulphide etc

N version - display range 100ppm to 50%

L version 0-1ppm to 5% (contains electrical solenoid valves to keep cell stored under inert gas)

H version 0.1ppm to 10% (suitable for use with hydrogen containing gases only - contains electrical solenoid valves to keep cell stored under inert gas)

Display resolution

 

 

0.1% from 10% to 100%

0.01% from 0.50% to 9.99%

10ppm from 500ppm to 4999ppm

1ppm from 50ppm to 499ppm

Stability

Better than 2% of full-scale per month

Cell life

E version - up to 5 years

N, L & H versions - up to 2 years

Sample flow

Between 100 and 500ml/min for optimum operation. Max 1 litre/mm.

Sample temperature

-5 to +40°C (non condensing)

Sample connections

Inlet and outlet: captive-seal compression fittings suitable for 0.25 inch (or 6mm) o/d tube

Speed of response

T90 - variable depending on sensor and concentration. Approximately 3s at % levels and 20s at ppm levels. The ppm figure assumes that the sensor is purged down to the ppm concentration of interest.

Analogue outputs

0 to 5V or 4 to 20mA (specify at time of order) Each user-programmable to the following spans:

   E version 0 to 100%, 0 to 25%, 0 to 5%

   N version 0 to 100%, 0 to 25%, 0 to 5%, 0 to 5000ppm, 0 to 500ppm

   L & H version 0 to 5%, 0-5000ppm, 0-500ppm, 0-50ppm

Alarm outputs

 

Two alarms, each user-programmable for mode - HIGH, LOW

or OFF Hysteresis programmable from 0 to 10% of set point Volt-free changeover contacts rated at 250V r.m.s.a.c max., 3A max., max power 60W, normally energised.

Ambient temperature

 

0 to 40°C (continuous operation)

-5 to +55°C (intermittent operation)

Up Next

 

Cambridge Sensotec Limited
Unit 8, Royce Court
Burrel Road

St. Ives
CAMBS
PE27 3NE
ENGLAND

Tel. +44 (0)1480 462142
  Fax. +44 (0)1480 466032
mobile (07866) 624236
  sales@cambridge-sensotec.co.uk
Registered in England No. 4035592
VAT No. 745 9399 75

List of Gases we supply detectors for: oxygen (O2), carbon dioxide (CO2), carbon monoxide (CO), hydrogen sulfide (H2S), Explosive gas (% LEL), hydrogen (H2), helium (He), argon (Ar), neon (Ne), Freon, nitrogen (N2), ammonia (NH3), hydrogen chloride (HCl), hydrogen cyanide (HCN), ozone (O3) hydrogen fluoride (HF), nitric oxide (NO), nitrogen dioxide (NO2), ethylene (C2H4), chlorine (Cl2), ethylene oxide (ETO), sulfur dioxide (SO2), silane (SiH4), phosphine (PH3), arsine (ASH3), chlorine dioxide (CLO2), phosgene (COCl2), fluorine (F2), hydrazine (N2H4), volatile organic components (VOC), acetone ( CH3COCH3), acetylene (C2H2), benzene (C6H6), butane (C4H10), methane (CH4), pentane (C5H12), propane (C3H8), toluene (C7H8), xylene (C8H10), gasoline, diesel & kerosene CLICK HERE FOR SITE MAP