Rapidox 1100 Technical Characteristics

Analyzer dimension

250mm X 263mm X 150mm

Weight

3 kg

Power supply

110/120V or 220/240V, 50/60Hz; Max power consumption 4VA

Display

Multi-digit LCD - character height 12.7mm

Display ranges

E version - display range 0.01% to 100.0%. Suitable for samples containing mildly acidic gases; eg, carbon dioxide, hydrogen sulphide etc, hydrogen & helium

 

L version - display range 0.1ppm to 10,000ppm (1%).  Suitable for low ppm measurements in hydrogen, helium and gases containing solvents, VOCs and fuels

 

Z version - display rang 1ppm to 100.0%.  Suitable for inert gases such as argon and nitrogen

Accuracy

+/- 1% of full scale for the E & L version

+/- 1% per decade for the Z version

Stability

Better than 2% of full-scale per month

Cell life

Up to 5 years (2 years for the L version)

Sample flow

1.2 litre/min flow rate (flow can be adjusted using the menu system

Sample temperature

-5 to +40°C (non condensing)

Sample connections

Inlet and outlet: captive-seal compression fittings suitable for 0.25 inch (or 6mm) o/d tube
L version has self sealing coupling on gas inlet

Speed of response

T90 - variable depending on sensor and concentration. Approximately 3s at % levels and 20s at ppm levels. The ppm figure assumes that the sensor is purged down to the ppm concentration of interest.

The L version 180 seconds from 10,000 ppm down to 1000ppm and <5 hours from 10ppm to 1ppm

Analogue outputs

0 to 10V and 4 to 20mA (specify at time of order) Each user-programmable range

Alarm outputs

Two alarms, each user-programmable for mode - HIGH, LOW

Ambient temperature

 

0 to 40°C (continuous operation)

-5 to +55°C (intermittent operation)

Up Next

 

Cambridge Sensotec Limited
Unit 8, Royce Court
Burrel Road

St. Ives
CAMBS
PE27 3NE
ENGLAND

Tel. +44 (0)1480 462142
  Fax. +44 (0)1480 466032
mobile (07866) 624236
  sales@cambridge-sensotec.co.uk
Registered in England No. 4035592
VAT No. 745 9399 75

List of Gases we supply detectors for: oxygen (O2), carbon dioxide (CO2), carbon monoxide (CO), hydrogen sulfide (H2S), Explosive gas (% LEL), hydrogen (H2), helium (He), argon (Ar), neon (Ne), Freon, nitrogen (N2), ammonia (NH3), hydrogen chloride (HCl), hydrogen cyanide (HCN), ozone (O3) hydrogen fluoride (HF), nitric oxide (NO), nitrogen dioxide (NO2), ethylene (C2H4), chlorine (Cl2), ethylene oxide (ETO), sulfur dioxide (SO2), silane (SiH4), phosphine (PH3), arsine (ASH3), chlorine dioxide (CLO2), phosgene (COCl2), fluorine (F2), hydrazine (N2H4), volatile organic components (VOC), acetone ( CH3COCH3), acetylene (C2H2), benzene (C6H6), butane (C4H10), methane (CH4), pentane (C5H12), propane (C3H8), toluene (C7H8), xylene (C8H10), gasoline, diesel & kerosene CLICK HERE FOR SITE MAP